Another member among eralytics’ flash point testers is the eraflash lt. It is based on the unrivalled performance of eraflash and adds new benefits for fast low-temperature flash point testing. The stacked Peltier design allows measurements in full compliance with ASTM D7094, ISO 24966 and IP620 with oven temperatures as low as -40 °C. The cooling rates are much faster than in any other instrument on the market.
eraflash lt at a glance
360° View
Low temperature flash point testing
To increase sample throughput for low-temperature flash point testing, eraflash lt is equipped with stacked Peltier elements. This provides three major improvements for low-temperature testing:
- Faster turnaround: Cooling speed is increased by approximately 35%.
- Expanded stand-alone range: Without external cooling, temperatures approximately 35 °C below ambient (e.g., -15 °C) are easily accessible.
- Maximum low-temperature capability: With external water cooling, low-temperature flash point testing down to an oven temperature of -40 °C is possible.
Simple, low-temperature flash point testing is now achievable without the need to cool the entire instrument or environment prior to measurement.

Further improvements for low temperature flash point testing
eraflash lt is equipped with a coolant valve that automatically shuts off coolant flow when not needed. This helps eliminate humidity condensation on the cooler. During cool-down, ice formation on the oven surface is prevented by using a cooling cup that shields the oven from ambient air. eraflash lt also allows samples to be pre-cooled prior to the actual measurement.

Applications
eraflash lt is the ideal partner when measuring samples with flash points below ambient temperature or even below 0 °C. According to ASTM D7094 and IP 620, flash point measurement must start 18 °C below the expected flash point. For a sample like ethanol with an expected flash point of around 12 °C, the measurement must start at -6 °C. eraflash lt speeds up the analysis because it is optimized for low-temperature flash point measurements. Flash points of crude oils, especially those containing highly volatile compounds, can fall well below 0 °C. Industries with high sample throughputs, such as the fragrance industry, also benefit from fast turnaround times due to the instrument’s enhanced cooling rate.

Measurement procedure
Just four easy steps are required for safe, low-temperature flash point testing at maximum precision:
- Enter the sample ID and operator name
- Select the measurement standard and add additional parameters (e.g., expected flash point)
- Fill the sample cup with sample (1 mL or 2 mL), place the sample cup inside the measurement chamber, and close the door
- Press the RUN button
The instrument now visualizes the measurement procedure on the screen, showing the sample and oven temperatures, as well as a pressure gauge displaying the pressure increase after ignition. After the measurement, the flash point temperature is automatically corrected for atmospheric pressure and rounded to the nearest 0.5 °C. The result is saved in the corresponding result file. To improve the measurement further, a cooling cup can be used to prevent ice formation during the cool-down process.
Specific characteristics of the low-temperature flash point test can be observed using Combustion Graphics™, making eraflash lt the ideal solution for both R&D and quality control.
scroll to see more images
gallery controls


eraflash lt – Low-temperature measurements
Take a look at our latest video that demonstrates how easy low-temperature flash point testing can be done using eraflash or the flashpoint autosampler tester eraflash s10.
eraflash lt in detail
Technical Specifications for eraflash lt
- Standard Test Methods
- ASTM D6450, D7094, D8606, ISO 24966, IP 620, SH/T 0768
- Fuel Specifications (ASTM D7094)
- ASTM D396, D975, D1655, D2880, D3699, D7467, D7566
- Approved by
- NATO, RCRA, US DOT, US NAVY
- Special Test Methods
- Fast screening test method SCAN for unknown samples, Flash / No Flash tests, freely programmable test methods
- Correlation Test Methods
- Pensky-Martens Closed Cup: ASTM D93; EN ISO 2719; DIN 51758; IP34; JIS K2265
TAG Closed Cup: ASTM D56
Abel-Pensky Closed Cup: ISO 13736; IP170
Small Scale Closed Cup: ASTM D3828; EN ISO 3679, EN ISO 3680 - Stacked Peltier Technology
- High-speed measurements at mid and low temperatures
-40 °C to 120 °C (-40 °F to 248 °F) - CPT – Contamination Prevention Technology™
- Advanced electrode protection and self-cleaning ignition system to minimize cleaning and maintenance
- Combustion Graphics™
- Display of combustion characteristics for contamination analysis
- QuickCal™
- In-situ calibration with eralytics unique calibration cups without the need of dismantling the instrument
- Fuel Dilution Program
- Automatic fuel dilution measurement for the analysis of used engine oils
- Temperature Range
- -15 °C to 120 °C (5 °F to 248 °F) stand alone down to -40 °C (-40 °F) with external cooling
- Temperature Stability
- 0.1 °C (0.2 °F)
- Sample Troughput
- Up to 12 samples per hour
- Sample Volume
- 1 mL (ASTM D6450), 2 mL (ASTM D7094, ISO 24966, IP620)
- Sample Cups
- Nickel-plated aluminum, stainless steel cups available upon request
- Result Database
- Over 100 000 detailed test reports stored in internal memory
- Alarm Tracking
- All alarm messages are stored in the database together with the result
- Display
- Industry proven 8.4″ multilingual color touchscreen
- Languages
- English, German, French, Spanish, Russian, Chinese
Additional langages available upon request - Interfaces
- Built-in PC with Ethernet, USB and RS232 interfaces
Wi-Fi via external USB dongle
Direct LIMS connectivity and output to printer or PC
Optional input by external keyboard, mouse and barcode / QR-code reader - Remote Control
- Remote service capability via Ethernet interface
- PC software
- erasoft RCS – remote control Windows® software for multi-instrument remote control, convenient data transfer andresult analysis
- Result database
- Over 100 000 detailed test reports stored in internal memory
- Result database
- All alarm messages are stored in the database together with the results
- Operating Conditions
- Temperature range 10 °C to 40 °C
Humidity up to 90% RH, non-condensing - Power Requirements
- Auto-switching 85-264 V AC, 47-63 Hz, 70 W (multi-voltage power supply)
Field application: 12 V/9 A DC (vehicle battery) - Dimensions & Weight
- (W x D x H) = 24.4 x 32.9 x 34.8 cm (9.6 x 12.9 x 13.7 in) / 9.9 kg (21.8 lb)