eraflash lt

The low-temperature flash point tester.

Another member among eralytics’ flash point testers is the eraflash lt. It is based on the unrivalled performance of eraflash and adds new benefits for fast low-temperature flash point testing. The stacked Peltier design allows measurements in full compliance with ASTM D7094 and IP620 with oven temperatures as low as -40° C. The cooling rates are much faster than in any other instrument on the market.

360° View

Low temperature flash point testing

To increase sample throughput for low temperature flash point testing the eraflash lt was equipped with stacked Peltier elements. This leads to three major improvements for low temperature flash point testing:

  • The cooling speed of the instrument was increased by approximately 35%
  • Without any external cooling temperatures approximately 35 °C below ambient temperature (e.g -15 °C) are accessible.
  • With external water cooling low temperature flash point testing down to -40 °C oven temperature is possible

Now simple low temperature flash point testing is possible without the need to cool the equipment with dry ice or other substances.

Further improvements for low temperature flash point testing

eraflash lt is equipped with a coolant valve that automatically shuts the coolant flow off if not needed. This helps to eliminate condensation of humidity on the cooler. Ice formation during cool down on the oven surface is prevented by using a cooling cup that shields the oven surface from the ambient air. It is also possible to cool the sample using eraflash lt prior to the actual measurement.

Applications

eraflash lt is the ideal partner if samples with flash points below ambient temperature or even below 0 °C are measured. According to ASTM D7094 and IP620 the flash point measurement needs to be started 18 °C below the expected flash point. For a sample like Ethanol with an expected flash point of around 12 °C the measurement needs to be started at -6 °C. Using eraflash lt speeds up the analysis as it is optimized for low temperature flash point measurements. Flash points of crude oils, especially if they still contain volatile compounds, can reach flash points well below 0 °C. Industries with high sample throughputs such as the fragrances industry also profit from the fast turnaround time due to the increased cooling rate.

Bottles of different fragrances

Measurement procedure

Just four easy steps are required for a safe low temperature flash point testing at maximum precision:

  • Enter sample ID and operator name
  • Select the measurement standard and add additional parameters (e.g. expected flash point)
  • Fill the sample cup with sample (1 mL or 2 mL), place the sample cup inside the measurement chamber, and close the door
  • Press the RUN button

The instrument now visualizes the measurement procedure on the screen showing the sample and oven temperature as well as pressure gauge displaying the pressure increase after ignition. After the measurement the flash point temperature is corrected for the atmospheric pressure and rounded to the next 0.5 °C. The result will be saved in the corresponding result file. To improve the measurement further, it is possible to use a cooling cup to prevent ice formation during the cool down process.

Characteristics of the low temperature flash point testing can be seen in the Combustion Graphics™. This makes eraflash lt the ideal solution for R&D as well as quality control.

gallery gallery controls
An ERAFLASH flah point tester running a measurement and visualizing the measurement process on the screen
An ERAFLASH flash point tester showing the final result of a measurement on the screen

eraflash lt – Low-temperature measurements

Take a look at our latest video that demonstrates how easy low-temperature flash point testing can be done using eraflash or the flashpoint autosampler tester eraflash s10.

You are currently viewing a placeholder content from YouTube. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.

More Information

Technical Specifications

Available Test Methods
ASTM D6450, ASTM D7094; IP620; SH/T 0768
Pre-Programmed Correlation Methods to…
Pensky Martens Closed Cup: ASTM D93, EN ISO 2719, DIN 51758, IP34, JIS K2265
TAG Closed Cup: ASTM D56
Abel Pensky Closed Cup: ISO 13736, IP170
Small Scale Closed Cup and Flash / No Flash methods: EN ISO 3679, EN ISO 3680, ASTM D3828
Fuel Specifications (ASTM D7094)
ASTM D396, ASTM D975, ASTM D2880, ASTM D3699, ASTM D7467
Speed Tests
Fast screening test programs for unknown samples
Stacked Peltier Technology
High speed measurements at mid and low temperatures
-40 °C to 120 °C (-40 °F to 248 °F)
CPT – Contamination Prevention Technology™
Advanced electrode protection and self-cleaning ignition system to minimize cleaning and maintenance
Combustion Graphics™
Display of combustion characteristics for contamination analysis
QuickCal™
In-situ calibration with eralytics unique calibration cups without the need of dismantling the instrument