eraflash s10

The safest flash point autosampler.

eralytics’ flash point autosampler eraflash s10 combines the safest measurement with a 10-position autosampler. The inherently safe CCCFP methods ASTM D7094, ASTM D6450, and IP620 are ideal for unattended measurements. ERAFLASH S10 ignites the vapors of 1-2 ml of the sample inside a closed chamber and measures according to the pressure increase.

360° View

Autosampler for flash point testing

eraflash s10 is the first 10-position autosampler for flash point testing. It uses the inherently safe methods ASTM D7094, ASTM D6450, and IP620. With a low sample volume of only 1-2 ml and the continuously closed sample cups, they are ideally suited to build a completely safe autosampler for flash point testing. eraflash s10 is the perfect solution for high-throughput laboratories and takes up very little laboratory space.

This makes ERAFLASH S10 the ideal analyzer for unattended flash point testing.

 

Sample cup carousel of the ERAFLASH S10 with the 10th cup missing showing the engraved sample number

Cost-efficient and fast flash point testing

When looking at valuable products, especially with small batch sizes, flash point testing using the well-known Pensky Martens principle can consume a significant amount of sample, thus leading to high analytical costs. When performing flash point testing according to ASTM D7094 and IP620  just 2 ml of the sample are needed. ASTM D6450 reduces this volume even further to 1 ml.

eraflash s10 is therefore a welcome addition to testing labs handling large amounts of costly samples every day. This includes fragrances, paints, and varnishes producers.

This image shows the two different sample cups for ASTM D6450 and ASTM D7094

Peltier Boost Technology®

Similar to eralytics’ single position flash point tester the patented PBT – Peltier Boost Technology® allows flash point testing below 0 °C (32 °F) and above 200 °C (392 °F) with a single instrument. PBT separates the Peltier elements during the heat up at 100 °C shielding them from higher temperatures that would normally harm them. During the cool down, however, they can be used for active cooling starting at 200 °C and thus reduce the turnaround time significantly. PBT also facilitates the same temperature range of -25 °C (-13 °F) to 420 °C (788 °F) of our single position flash point tester in an autosampler for flash point testing.

gallery gallery controls
Flash Point Measurment Screen
Result screen when measuring Viscogen and n-Heptan with the 10-position autosampler ERAFLASH S10 according to ASTM D7094

A cooling lid for shorter turnaround times

eraflash s10’s optional cooling lid is an unsurpassed benefit, especially in applications that involve highly volatile samples. The lid keeps the pending samples at a stable temperature, which makes volatile compounds far less likely to evaporate before their measurement starts. Therefore, results are more precise and more reliable. With pre-cooled sample cups cooling during the measurement can be less intense, which results in shorter turnaround times.

Measuring Graph showing results with/without ERAFLASH S10 Cooling Lid

Measurement procedure

Just four easy steps are required for an unattended safe and fast flash point testing at maximum precision:

  • Enter sample IDs and operator name
  • Select the measurement standards and add additional parameters (e.g. expected flash point)
  • Fill up to 10 sample cups with samples, place the sample cups on the carousel and close the cover plate
  • Press the RUN button

During flash point testing the instrument visualizes the measurement procedure on the screen showing the sample and oven temperature as well as pressure gauge displaying the pressure increase after ignition. Additional information about the currently measured sample, as well as finished and still samples to be measured are displayed. After the measurements the flash point temperature is corrected for the atmospheric pressure and rounded to the next 0.5°C. All results will be saved in the corresponding result file.

gallery gallery controls
ERAFLASH S10 before measurement
ERAFLASH S10 Sample Selection
ERAFLASH S10 Measuring
ERAFLASH S10 Results
ERAFLASH S10 result details

eraflash s10 feature video

  • High throughput with the 10-position autosampler
  • Inherently safe with no open flame
  • Only a very low sample volume of 1-2 mL needed
  • Takes up very little laboratory bench space
  • Fully compliant with ASTM D6450, ASTM D7094, and IP620
  • Measurement of fuels, lubricants, fragrances, flavors, paints, …

You are currently viewing a placeholder content from YouTube. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.

More Information

eraflash s10 operation video

Watch this video to get an overview on the standard operation of eraflash s10: From installation to measurement and some extended measuring options.

You are currently viewing a placeholder content from YouTube. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.

More Information

Technical Specifications

Available Test Methods
ASTM D6450, ASTM D7094, IP620, SH/T 0768
Pre-Programmed Correlation Methods to…
Pensky Martens Closed Cup: ASTM D93, EN ISO 2719, DIN 51758, IP34, JIS K2265
TAG Closed Cup: ASTM D56
Abel Pensky Closed Cup: ISO 13736, IP170
Small Scale Closed Cup and Flash / No Flash methods: EN ISO 3679, EN ISO 3680, ASTM D3828
Fuel Specifications (ASTM D7094)
ASTM D396, ASTM D975, ASTM D2880, ASTM D3699, ASTM D7467
Speed Tests
Fast screening test programs for unknown samples
Sample Handling
10 position autosampler
PBT – Peltier Boost Technology®
High speed heating & cooling
-25 °C to 420 °C (-13 °F to 788 °F) with a single analyzer
CPT – Contamination Prevention Technology™
Advanced electrode protection and self-cleaning ignition system to minimize cleaning and maintenance
Combustion Graphics™
Display of combustion characteristics for contamination analysis
QC Mode
Built-in quality control mode including on-screen QC charts
Cooling Lid Temperature
5 °C to 30 °C (optional)
QuickCal™
In-situ calibration with eralytics unique calibration cups without the need of dismantling the instrument
Temperature Range
0 °C to 200 °C (32 °F to 390 °F) stand alone
down to -25 °C (-13 °F) with external cooling
up to 420 °C (788 °F) with optional High Temperature Module
Operating Conditions
Temperature range 10 °C to 40 °C
Humidity up to 90% RH, non-condensing
Power Requirements
Auto-switching 85-264 V AC, 47-63 Hz, max. 150 W (multi-voltage power supply)
Field application: 12 V DC (vehicle battery) adapter available